The Triple Plus+ is the latest generation of the respected “Triple” family of toxic, oxygen and flammable gas detectors. With over 90,000 units sold to oil and gas, chemical, utility and manufacturing companies worldwide, their combination of ruggedness, reliability, ease of use and sensor flexibility has set the standard for multigas portables.
IR option IR sensors have very high resistance to interference or inhibition by other gases and are suitable for both high gas concentrations and use in inert (oxygen free) backgrounds where catalytic pellistor sensors would perform poorly.
The unique design of the Crowcon IR sensor module used in Triple Plus+ IR allows measurement of hydrocarbons at both %LEL and % volume ranges making this instrument ideal for tank and line purging applications.
No poisoning effects – works well where catalytic pellistors would perform poorly
Long life sensor and no burnout – reduced cost of ownership
% vol. measurement for purging applications – offers flexibility
Selection of correction factors for flammable gas – accurate detection of a range of Hydrocarbons
Typical sensors life; Combustible catalytic 5 years, IR 5 years, Toxic 3 years and Oxygen 2 years
Vivid LED alarms with top visibility and audible 85 dB
Confidence blip, low battery and TWA warnings
Timed data logging with 100-hour capacity, maximum 6000 data points per channel and events
User and site names logging
LCD with backlight four gases displayed simultaneously at current, TWA or peak levels
IR Flammable, single or dual-range measurement
Ingress protection IP65
Operating temperature -20°C to +50°C
Buoyant recoverable from water
Rechargeable battery run time, standard instrument >12 hours, pumped >10 hours
Size 118 x 200 x 73mm, weight 880g
Size: 118 x 200 x 73mm (4.65 x 7.8 x 2.9ins)
Weight: 880g (31oz)
Approval codes: Europe: ATEX II 2G: Ex ib d IIC T4 Gb (Ta = -20°C to +50°C); North America: Class I Division 1, Groups A, B, C and D; IECEx Ex ib d IIC T4, (Ta = -20°C to +50°C) SIR06.0084X
Safety standards: Europe: EN50014, EN50020, EN50018-2, EN50270, EN50082-2; North America: UL 913, CSA 22.2 No.157 (SFA3009 Triple Plus+)
Ingress protection: IP65
Typical response times: Methane 20s (Triple Plus+) IR 20s (Triple Plus+IR), Oxygen 10s, typical Toxic Gas 20s
Expected sensor life: Flammable: 5 years (Triple Plus+), IR: 5 years (Triple Plus+ IR) Toxic: 3 years (typical), Oxygen: 2 years
Audible alarms: Piercing, rapid intermittent, piezo-electric tone 85 dB at 1 m; Confidence blip (selectable); Distinctive double blip for low battery with user accept
Visible alarms: Gas hazard: vivid flashing red LED; Instrument is working: slow flashing green LED; Low battery: distinctive double flash of green LED
Display:
2 lines of 24 characters, dot matrix high contrast; LCD with back light operating for 30 minutes
Values and units for up to 4 gases displayed simultaneously at current, TWA or peak levels
Low battery, pump failure, real time and elapsed time, calibration and configuration data also displayed
Data logging: Adjustable sample rate, 100 hours at 1-minute sample rate (maximum 6000 data points per channel), logging of events, user and site
LIBRA:
Li-Ion rechargeable battery pack
Battery Life: standard instrument >12 hours
Battery Life: pumped instrument >10 hours
Battery:
Charging: Dual rate drop-in charger;
Dual rate charger/PC interface
Rechargeable Li-ion batterypack
Standard instrument >12 hours
Pumped instrument >10 hours
Sampling: Optional internal pump with on/off button
Operating temperature range: -20°C to +50°C (-4°F to 122°F)*
Humidity: 20 to 90% RH non-condensing**
PC Interface:
RS232 connection via the interface charger
RS232/USB converter available
Portables PC also controls data logging, giving the user a choice of direct graphical
output or a data file compatible with all leading spreadsheets
Calibration: Triple Plus+ can be routinely calibrated using Portables PC software
Accessories:
Straight and telescopic probes, water trap, baffle plate, aspirator, calibration flow adaptor and hose
Leather carry case***, is sold as an integral part of the unit and is not an accessory.
Standard: Manual, shoulder strap and allen keys (pumped units include additional flow plate, 2m hose and pouch)
* Toxic and oxygen sensors are not rated for continuous operation at high temperatures.
Consult Crowcon if operating temperatures are likely to exceed 40°C (104°F) for extended periods.
** Toxic and oxygen sensors RH specifications vary, consult Crowcon for specific operating humidities.
*** Leather carry case is sold as an integral part of the unit and is not an accessory. Triple Plus+ must be used within the Leather gas to meet certification requirements.
The Triple Plus+ has two dual rate charging solutions supplied with UK, EU, US plug and without plug configurations. A vehicle adapter for in-vehicle charging is also available Standard drop-in charging cradle, PC interface charger
(1) Manual
(1) Shoulder strap
(1) Allen keys (pumped units include additional flow plate, 2m hose and pouch)